SAHPRA recall of BiCOAG probes over risk of ceramic tip detachment

Two lots of the BiCOAG Hemostasis Probe are being recalled in South Africa after Olympus, the original equipment manufacturer, determined that affected models may have been manufactured without the designated amount of adhesive at the rear of the ceramic tip, which may result in the tip detaching.

Endoscopy units and theatre stores must stop using and distributing the affected lots immediately, check internal inventory and quarantine stock. The device is used for coagulation of active bleeding lesions in the upper or lower gastrointestinal tract during endoscopic procedures, so a detached ceramic tip would be released into the GI tract during a procedure being performed on a patient who is already bleeding. Units should also consider whether any procedures were performed using these lots.

LotExpiryFirst distributed
KR31375418/10/202615 February 2024
KR31373506/10/202616 January 2025

The recall is classified Class II Type C and dated 28 August 2026. It is being conducted by Surgical Innovations (Pty) Ltd of Randburg in consultation with the South African Health Products Regulatory Authority, on behalf of the manufacturer. Both lots are supplied as single units and expire in October 2026.

Source: https://www.sahpra.org.za/document/bicoag-hemostasis-probe/

About the author

Drug Eye Africa Newsroom

Drug Eye Africa Newsroom

The Drug Eye Africa newsroom reports on medicines regulation, recalls and medicine quality across the African continent. Alerts are matched against the issuing authority's published notice before publication, and corrections are timestamped on the article rather than made silently.

Regulatory alerts and news, direct to your inbox

Independent pharmaceutical news and regulatory intelligence covering Africa.

Scroll to Top